Certificate of Analysis · Lot release

B7-33 2 mg

Lot PL-2610-243 · Lyophilized powder · Label claim 2 mg per vial
Tested in-house by PrymaLab
In-house analytical result,
not an independent third-party test
Lot / batch number
PL-2610-243
Date made
2026-10-08
Date of analysis
2026-10-09
Date issued
2026-10-23
Retest / expiry
2028-10-08
Units in lot / sampled
183 / 2
Storage
—
Certificate type
Lot COA, 2 pages
Compound
B7-33
CAS number
1818415-56-3
Molecular formula
—
Theoretical mass (avg.)
2,987.60 Da
Sequence
VIKLSGRELVRAQIAISGMSTWSKRSL
Purity · RP-HPLC, area %
99.54%
Specification ≥ 99.0% • main peak at 16.41 min
Identity · Observed Mass
2,987.59 Da
Theoretical 2,987.60 Da • −0.01 Da (3 ppm)
Overall conclusion
PASS
Meets every specification for the tests performed
TestMethodSpecificationResultConclusion
Identity (mass)LC-MS, ESI positiveTheoretical ± 1.0 Da2,987.59 DaPASS
PurityRP-HPLC, UV 214 nm, area%≥ 99.0%99.54%PASS
Largest impurityRP-HPLC, area%≤ 1.0%0.23% (RT 17.70)PASS
Water contentKarl Fischer≤ 5.0%3.09%PASS
Bacterial endotoxinLAL, kinetic≤ 5.0 EU/mg0.377 EU/mgPASS
SterilityUSP <71>, 14-dayNo growthNo growthPASS

1HPLC chromatogram

Data file: PL-2610-243-HPLC.raw · Column C18, 150 × 4.6 mm · UV 214 nm
0.01%0.1%1%10%100%141516171819Retention time (min) · stick height = area % (log scale)99.54% · 16.41 min0.06% †0.17% †0.23% †
Peak plot drawn from the integrated peak table (area %, log scale); it is not the instrument trace, which is retained in PL-2610-243-HPLC.raw. † Nominal reference retention time for the assigned impurity; the main-peak RT is observed in this lot.
PrymaLab
PrymaLab
Certificate of Analysis · continued
B7-33 2 mg · Lot PL-2610-243 · Certificate PL-26-00356
Lot PL-2610-243
Method, peak table and approval

2HPLC method · PL-2610-243-HPLC.raw

Instrument
Waters Alliance e2695
Data System
Empower 3
Detector
UV absorbance
Detection Wavelength
214 nm
Column
C18, 150 × 4.6 mm, 3.5 µm
Mobile Phase A
0.1% TFA in water
Mobile Phase B
0.1% TFA in MeCN
Gradient
5–95% B over 25 min (linear)
Flow Rate
1.0 mL/min
Column Temperature
40 °C
Injection Volume
10 µL

3Gradient program

Time (min)% A% B
0.0955
25.0595

4Peak table

PeakRT (min)AreaArea %
115.00 †—0.06
215.80 †—0.17
316.41—99.54
417.70 †—0.23
Total—100.00
Peaks: 1 deamidation product; 2 truncated sequence; 3 main peak (B7-33); 4 oxidized variant · Area % from the integrated peak table (absolute areas are not part of the lot-release export) · † nominal reference RT for the assigned impurity · Purity = main peak area ÷ total integrated area × 100, to two decimals; main peak shaded.

5Identity by mass spectrometry

Instrument
Waters SQD
Ionization / Mode
ESI, positive
Reference m/z (charge)
996.87 (3+) · 747.91 (4+)
Deconvoluted Mass
2,987.59 Da
Theoretical Mass
2,987.60 Da
Difference
−0.01 Da (3 ppm)
Data File
PL-2610-243-MS.d
Acceptance
Within ± 1.0 Da

6Additional tests

TestPerformed byDateReport / file
Water content (Karl Fischer)PrymaLab Analytical2026-10-09—
Endotoxin (LAL, kinetic)PrymaLab Analytical2026-10-09—
Sterility (USP <71>, day-14 read-out)PrymaLab Analytical2026-10-23—
A test with no row here was not performed on this lot. Outside-laboratory reports, when used, are attached and named below.

7Review and approval

D. Reynolds
Analyst · performed the testing
Date: 2026-10-23
S. Calloway
Reviewer · checked against raw data
Date: 2026-10-23
M. Phelps
Lab Director · approved for release
Date: 2026-10-23