3705 Music Branch Rd, Ashland, KY 41102 · +1 (606) 465-9063 · prymalab.net
Certificate no.
PL-26-00186
Verification key
UKNTEP78
Scan or enter both at prymalab.net/testing/verify
Certificate of Analysis · Lot release
GHK-Cu 100 mg
Lot PL-2610-068 · Lyophilized powder · Label claim 100 mg per vial
Tested in-house by PrymaLab
In-house analytical result, not an independent third-party test
Lot / batch number
PL-2610-068
Date made
2026-10-02
Date of analysis
2026-10-05
Date issued
2026-10-19
Retest / expiry
2028-10-02
Units in lot / sampled
369 / 2
Storage
−20 °C, desiccated, dark
Certificate type
Lot COA, 2 pages
Compound
GHK-Cu
Copper tripeptide-1; Gly-His-Lys · Cu(II)
CAS number
89030-95-5
Molecular formula
C14H22CuN6O4
Theoretical mass (avg.)
401.91 Da
Sequence
GHK-Cu(II)
Purity · RP-HPLC, area %
99.71%
Specification ≥ 99.0% • main peak at 10.72 min
Identity · Observed Mass
401.93 Da
Theoretical 401.91 Da • +0.02 Da (50 ppm)
Overall conclusion
PASS
Meets every specification for the tests performed
Test
Method
Specification
Result
Conclusion
Identity (mass)
LC-MS, ESI positive
Theoretical ± 1.0 Da
401.93 Da
PASS
Purity
RP-HPLC, UV 214 nm, area%
≥ 99.0%
99.71%
PASS
Largest impurity
RP-HPLC, area%
≤ 1.0%
0.13% (RT 9.30)
PASS
Water content
Karl Fischer
≤ 5.0%
3.25%
PASS
Bacterial endotoxin
LAL, kinetic
≤ 5.0 EU/mg
0.245 EU/mg
PASS
Sterility
USP <71>, 14-day
No growth
No growth
PASS
1HPLC chromatogram
Data file: PL-2610-068-HPLC.raw · Column C18, 150 × 4.6 mm · UV 214 nm
Peak plot drawn from the integrated peak table (area %, log scale); it is not the instrument trace, which is retained in PL-2610-068-HPLC.raw. † Nominal reference retention time for the assigned impurity; the main-peak RT is observed in this lot.
PrymaLab
Certificate of Analysis · continued
GHK-Cu 100 mg · Lot PL-2610-068 · Certificate PL-26-00186
Lot PL-2610-068
Method, peak table and approval
2HPLC method · PL-2610-068-HPLC.raw
Instrument
Waters Alliance e2695
Data System
Empower 3
Detector
UV absorbance
Detection Wavelength
214 nm
Column
C18, 150 × 4.6 mm, 3.5 µm
Mobile Phase A
0.1% TFA in water
Mobile Phase B
0.1% TFA in MeCN
Gradient
5–95% B over 25 min (linear)
Flow Rate
1.0 mL/min
Column Temperature
40 °C
Injection Volume
10 µL
3Gradient program
Time (min)
% A
% B
0.0
95
5
25.0
5
95
4Peak table
Peak
RT (min)
Area
Area %
1
9.30 †
—
0.13
2
10.72
—
99.71
3
11.50 †
—
0.04
4
12.20 †
—
0.12
Total
—
100.00
Peaks: 1 free ghk (copper-free); 2 main peak (GHK-Cu); 3 deamidation product; 4 copper-loss product · Area % from the integrated peak table (absolute areas are not part of the lot-release export) · † nominal reference RT for the assigned impurity · Purity = main peak area ÷ total integrated area × 100, to two decimals; main peak shaded.
5Identity by mass spectrometry
Instrument
Waters SQD
Ionization / Mode
ESI, positive
Reference m/z (charge)
402.92 (M+H)+
Deconvoluted Mass
401.93 Da
Theoretical Mass
401.91 Da
Difference
+0.02 Da (50 ppm)
Data File
PL-2610-068-MS.d
Acceptance
Within ± 1.0 Da
6Additional tests
Test
Performed by
Date
Report / file
Water content (Karl Fischer)
PrymaLab Analytical
2026-10-05
—
Endotoxin (LAL, kinetic)
PrymaLab Analytical
2026-10-05
—
Sterility (USP <71>, day-14 read-out)
PrymaLab Analytical
2026-10-19
—
A test with no row here was not performed on this lot. Outside-laboratory reports, when used, are attached and named below.
7Review and approval
D. Reynolds Analyst · performed the testing Date: 2026-10-19
S. Calloway Reviewer · checked against raw data Date: 2026-10-19
M. Phelps Lab Director · approved for release Date: 2026-10-19