Certificate PL-26-00170 · CJC-1295 noDAC+Ipamorelin 15 mg Pen · lot PL-2610-048
VerifyCertificate page
PrymaLab
PrymaLab
PrymaLab Analytical · In-house laboratory
3705 Music Branch Rd, Ashland, KY 41102 · +1 (606) 465-9063 · prymalab.net
Certificate no.
PL-26-00170
Verification key
L8S6UU5X
Scan or enter both at
prymalab.net/testing/verify

Certificate of Analysis · Lot release

CJC-1295 noDAC+Ipamorelin 15 mg Pen

Lot PL-2610-048 · Solution · Label claim 15 mg/mL
Tested in-house by PrymaLab
In-house analytical result,
not an independent third-party test
Lot / batch number
PL-2610-048
Date made
2026-10-02
Date of analysis
2026-10-05
Date issued
2026-10-19
Retest / expiry
2028-10-02
Units in lot / sampled
255 / 2
Storage
—
Certificate type
Lot COA, 2 pages
Compound
CJC-1295 (no DAC) / Ipamorelin Blend
Modified GRF(1–29) + Ipamorelin co-lyophilized 1:1 blend
CAS number
863288-34-0 · 170851-70-4
Molecular formula
—
Theoretical mass (avg.)
3,367.90 · 711.85 Da (CJC-1295 noDAC / Ipamorelin)
Composition and sequences
CJC-1295 noDAC: 30-aa GHRH analog · Ipamorelin: Aib-His-D-2Nal-D-Phe-Lys-NH2
Purity · RP-HPLC, area %
99.45%
Lowest component (CJC-1295 noDAC) • specification ≥ 99.0% for every component
Identity · Components Confirmed
2 / 2
Components confirmed by LC-MS within ± 1.0 Da: 3,368.10 / 711.81 Da
Overall conclusion
PASS
Meets every specification for the tests performed
TestMethodSpecificationResultConclusion
CJC-1295 noDAC: purity · largest impurity · massRP-HPLC 214 nm (area%); LC-MS, ESI+≥ 99.0% · ≤ 1.0% · ± 1.0 Da99.45% · 0.34% · 3,368.10 DaPASS
Ipamorelin: purity · largest impurity · massRP-HPLC 214 nm (area%); LC-MS, ESI+≥ 99.0% · ≤ 1.0% · ± 1.0 Da99.79% · 0.08% · 711.81 DaPASS
Assay (total blend)Assay vs label claim14.25–15.75 mg/mL15.23 mg/mLPASS
Bacterial endotoxinLAL, kinetic≤ 5.0 EU/mg0.229 EU/mgPASS
SterilityUSP <71>, 14-dayNo growthNo growthPASS

1HPLC chromatogram

Data file: PL-2610-048-HPLC.raw · Column C18, 150 × 4.6 mm · UV 214 nm
0.01%0.1%1%10%100%131415161718192021Retention time (min) · stick height = area % (log scale)99.45% · 15.74 min99.79% · 18.10 min0.04% †0.34% †0.17% †0.07% †0.08% †0.06% †
Peak plot drawn from the integrated peak table (area %, log scale); it is not the instrument trace, which is retained in PL-2610-048-HPLC.raw. † Nominal reference retention time for the assigned impurity; the main-peak RT is observed in this lot. Area % is per component (each component main peak + its impurities = 100%).
PrymaLab
PrymaLab
Certificate of Analysis · continued
CJC-1295 noDAC+Ipamorelin 15 mg Pen · Lot PL-2610-048 · Certificate PL-26-00170
Lot PL-2610-048
Method, peak table and approval

2HPLC method · PL-2610-048-HPLC.raw

Instrument
Waters Alliance e2695
Data System
Empower 3
Detector
UV absorbance
Detection Wavelength
214 nm
Column
C18, 150 × 4.6 mm, 3.5 µm
Mobile Phase A
0.1% TFA in water
Mobile Phase B
0.1% TFA in MeCN
Gradient
5–95% B over 25 min (linear)
Flow Rate
1.0 mL/min
Column Temperature
40 °C
Injection Volume
10 µL

3Gradient program

Time (min)% A% B
0.0955
25.0595

4Peak table

PeakRT (min)AreaArea %
114.20 †—0.04
214.90 †—0.34
315.74—99.45
416.30 †—0.17
516.80 †—0.07
618.10—99.79
718.90 †—0.08
819.50 †—0.06
Total—100.00 per component
Main peaks: 3 CJC-1295 noDAC · 6 Ipamorelin; all other peaks are the assigned impurities of the neighbouring component · area % within each component (main + impurities = 100% per component) · † nominal reference RT · Purity = main peak area ÷ total integrated area × 100, to two decimals; main peak shaded.

5Identity by mass spectrometry

Instrument
Waters SQD
Ionization / Mode
ESI, positive
Deconvoluted Mass
3,368.10 / 711.81 Da
Theoretical Mass
3,367.90 / 711.85 Da
Order
CJC-1295 noDAC / Ipamorelin
Difference
+0.20 / −0.04 Da
Data File
PL-2610-048-MS.d
Acceptance
Within ± 1.0 Da (each component)

6Additional tests

TestPerformed byDateReport / file
Assay vs label claimPrymaLab Analytical2026-10-05—
Endotoxin (LAL, kinetic)PrymaLab Analytical2026-10-05—
Sterility (USP <71>, day-14 read-out)PrymaLab Analytical2026-10-19—
A test with no row here was not performed on this lot. Outside-laboratory reports, when used, are attached and named below.

7Review and approval

D. Reynolds
Analyst · performed the testing
Date: 2026-10-19
S. Calloway
Reviewer · checked against raw data
Date: 2026-10-19
M. Phelps
Lab Director · approved for release